Quality Control of CR Imaging Plates
E. Samei, Ph.D.
samei@duke.edu
Version 1.4, Jan. 25, 1999
The quality control of CR imaging plates requires an
x-ray source, a display device (soft- or hard-copy), and a CR reader previously
evaluated for satisfactory performance. It is recommended that this procedure
to be performed on all the imaging plates on a monthly basis and
whenever the number of small white speck artifacts in the images obtained from
an imaging plate is objectionable. Note
that too frequent cleaning of the phosphor screens or the use of non-approved
cleaning solutions can discolor the phosphor.
Equipment
Needed:
1. A calibrated x-ray source
2. A calibrated
hard/soft-copy display
3. A CR reader and ID
station
4. Screen cleaning solution
(for Agfa systems, Curix screen solution) and cloth
5. Lint-free gloves for
handling the screens
6. A permanent marker
7. Cleaning solution for the
table (alcohol)
8. A piece of cloth (towel
or sheet) for cleaning the table
Cleaning
the Screens
1. Collect and stack all the
cassettes to be cleaned by the CR ID station.
2. “Read” each cassette. Verify that the cassette ID (the number in
the upper left hand corner of the screen in the “READ” menu matches the
cassette label. If not or if the label is
erased, write the cassette ID on the label area of the cassette.
3. Open the cassette and verify
that the number written on the back of the screen matches that of the
cassette. If not or if the screen number
is erased, write the cassette ID on the back of the screen.
4. Repeat the steps 1-3 for all
the cassettes.
5. Erase all the cassettes.
6. Select a room that is free
of dust, is well lighted, and has a large table area. An x-ray room with a table-top would be
ideal.
7. Clean the table-top with a
cleaning solution (alcohol) and a cloth.
8. Put gloves on.
9. Take the screens out of the
cassettes in an orderly fashion. Put cassettes and screens next to each
other to avoid mixing. Be sure not
to touch the white sensitive surface of the screen (be extra careful if you
have long fingernails).
10. Visually inspect the
cassettes and the screens for physical defects. Look for chipping,
discoloration, bends, and cracks in the screens. Any defects other than dust and small
chippings at the edges should be recorded in the log table (you may want to
record the location of such defects in the log table.
11. Remove the dust and small
particles on the screen by gently wiping the surface with the screen cleaning
cloth.
12. If dirt resists cleaning,
wipe it off using the cloth damped with the Curix cleaning liquid using a
circular or straight motion. AVOID USING
EXCESSIVE PRESSURE. Also avoid excessive
use of the cleaning solution.
13. Let the screen dry.
14. Clean the inside of the
cassette with the screen cleaning solution and cloth, specially the tube side.
15. Let the cassette dry.
16. Put the screen back into the
cassette. The screen should be
thoroughly dried without any residues before insertion back into the cassette.
17. Proceed to the next step.
Uniformity
Test
1. Set the exposure set up for
cross-table exposures of the CR cassettes using 80 kVp, 5 mAs, and 180 cm
source-to-image distance (SID). Make
sure the entire cassette area will be exposed.
2. Expose each cassette.
3. Identify the cassette in the
ID station under “Uniformity Test.” (Any arbitrary Accession number may be used
since the images will be evaluated on the PS5000 workstation and will not be
sent to PACS.) Use the following reading parameters:
Menu: SYSTEM DIAGNOSIS, FLAT FIELD
Speed Class: 200
4. Read the cassette in the CR
reader.
5. Evaluate the image on the
PS5000 Processing Workstation. The
image should be uniform without any artifacts. If the screen is not
properly cleaned, some white specks can be seen. The position of such artifacts
vary from screen to screen, and they should be eliminated by repeating the
cleaning steps outlined above.
6. Any artifacts other than
white specks are considered abnormal and should be recorded in the QC log. Gross
nonunifomities not removable by repeated cleaning should be brought to the
attention of the PACS Physicist and the tech supervisor, and the screen should
be removed from circulation.
7. Click the “INFO” icon. Under the “Exposure Info” section that is
listed in the window that opens up, read the value for the “Exposure Level(s)”
and record it in the QC log sheet.
8. Delete the Uniformity test
images from the PS5000 workstation after you are finished will all
the cassettes.
Some QC
Precautions
·
Too frequent cleaning of the phosphor screens can discolor the
phosphor.
·
The imaging plate that have not been used for more than 24 hours should
be erased before use.
CR Screens
QC Log
Cassette Size
|
Cassette No.
|
Date
|
Visual
Defects
|
Cleaning (Ö)
|
Nonuni-formities
|
Exposure Level
|
Comments
|
14 x 17
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
10 x 12
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
8 x 10
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
|||||||
``
|
No comments:
Post a Comment